Intraoral dental x-ray radiography in Bosnia and Herzegovina: study for revising diagnostic reference level value

Praskalo, Jovica; Beganović, Adnan; Milanović, Jovana; Stanković, Koviljka

(Oxford Univ Press, Oxford, 2020)

TY  - JOUR
AU  - Praskalo, Jovica
AU  - Beganović, Adnan
AU  - Milanović, Jovana
AU  - Stanković, Koviljka
PY  - 2020
UR  - http://TechnoRep.tmf.bg.ac.rs/handle/123456789/4433
AB  - This paper presents the study that would allow a revision of the currently valid diagnostic reference level in intraoral dental X-ray radiography in Bosnia and Herzegovina. The study was conducted for six procedures that are used to capture incisor, premolar and molar teeth, of both maxilla and mandible, in adults. Measurements were performed on 41 intraoral X-ray devices, 20 of which were systems with digital image receptor and 21 film-based X-ray systems. In this study, incident air kerma (K-i) and air kerma-area product (P-KA) were used as patient dose descriptors. The third quartile (Q(3)) of the measured incident air-kerma values was used to compare with the current national diagnostic reference level (DRL) for intraoral dental procedures. The obtained results of Q(3) for both types of devices, with film-based and digital image receptors, have shown that the third quartile values are lower (3.5 and 1.2 mGy, respectively) than the current national DRL (7.0 mGy). Hence, new data can be used to re-establish the diagnostic reference levels in intraoral dental X-ray examinations in Bosnia and Herzegovina.
PB  - Oxford Univ Press, Oxford
T2  - Radiation Protection Dosimetry
T1  - Intraoral dental x-ray radiography in Bosnia and Herzegovina: study for revising diagnostic reference level value
EP  - 99
IS  - 1
SP  - 90
VL  - 190
DO  - 10.1093/rpd/ncaa081
ER  - 
@article{
author = "Praskalo, Jovica and Beganović, Adnan and Milanović, Jovana and Stanković, Koviljka",
year = "2020",
abstract = "This paper presents the study that would allow a revision of the currently valid diagnostic reference level in intraoral dental X-ray radiography in Bosnia and Herzegovina. The study was conducted for six procedures that are used to capture incisor, premolar and molar teeth, of both maxilla and mandible, in adults. Measurements were performed on 41 intraoral X-ray devices, 20 of which were systems with digital image receptor and 21 film-based X-ray systems. In this study, incident air kerma (K-i) and air kerma-area product (P-KA) were used as patient dose descriptors. The third quartile (Q(3)) of the measured incident air-kerma values was used to compare with the current national diagnostic reference level (DRL) for intraoral dental procedures. The obtained results of Q(3) for both types of devices, with film-based and digital image receptors, have shown that the third quartile values are lower (3.5 and 1.2 mGy, respectively) than the current national DRL (7.0 mGy). Hence, new data can be used to re-establish the diagnostic reference levels in intraoral dental X-ray examinations in Bosnia and Herzegovina.",
publisher = "Oxford Univ Press, Oxford",
journal = "Radiation Protection Dosimetry",
title = "Intraoral dental x-ray radiography in Bosnia and Herzegovina: study for revising diagnostic reference level value",
pages = "99-90",
number = "1",
volume = "190",
doi = "10.1093/rpd/ncaa081"
}
Praskalo, J., Beganović, A., Milanović, J.,& Stanković, K.. (2020). Intraoral dental x-ray radiography in Bosnia and Herzegovina: study for revising diagnostic reference level value. in Radiation Protection Dosimetry
Oxford Univ Press, Oxford., 190(1), 90-99.
https://doi.org/10.1093/rpd/ncaa081
Praskalo J, Beganović A, Milanović J, Stanković K. Intraoral dental x-ray radiography in Bosnia and Herzegovina: study for revising diagnostic reference level value. in Radiation Protection Dosimetry. 2020;190(1):90-99.
doi:10.1093/rpd/ncaa081 .
Praskalo, Jovica, Beganović, Adnan, Milanović, Jovana, Stanković, Koviljka, "Intraoral dental x-ray radiography in Bosnia and Herzegovina: study for revising diagnostic reference level value" in Radiation Protection Dosimetry, 190, no. 1 (2020):90-99,
https://doi.org/10.1093/rpd/ncaa081 . .
1
1