Microstructure and metallic ion release of pure titanium and Ti-13Nb-13Zr alloy processed by high pressure torsion

Authorized Users Only
2016
Authors
Dimić, Ivana
Cvijović-Alagić, Ivana
Voelker, Bernhard
Hohenwarter, Anton
Pippan, Reinhard
Veljović, Đorđe
Rakin, Marko
Bugarski, Branko
Article (Published version)
Metadata
Show full item record
Abstract
Significant enhancement of mechanical properties of metallic biomaterials can be achieved by grain refinement obtained by severe plastic deformation. The purpose of this study was to determine metallic ion release from commercially pure titanium (CPTi) and Ti-13Nb-13Zr alloy processed by high pressure torsion (HPT). The materials microstructures, in the initial state and after HPT deformation, were examined by scanning and transmission electron microscopy. The microhardness was determined along the radius of the disc-shaped samples of ultrafine-grained (UFG) CPTi and Ti-13Nb-13Zr alloy in order to evaluate homogeneity of HPT-processed materials. The quantities of released ions were determined using inductively coupled plasma-mass spectrophotometer for samples immersed in artificial saliva at 37 degrees C for 7 days. Also, the effect of artificial saliva pH value on metallic ion release was estimated. Obtained results revealed that the quantities of released ions from UFG CPTi and Ti-13...Nb-13Zr alloy obtained by HPT process were higher than the quantities of released ions from CPTi and Ti-13Nb-13Zr alloy produced by traditional casting. This behavior can be explained by the fact that metallic ions are easily released from microstructure with smaller grains achieved by HPT process.

Keywords:
High pressure torsion / Ultrafine-grained / Titanium / Ti-13Nb-13Zr alloy / Microhardness / Metallic ion release
Source:
Materials & Design, 2016, 91, 340-347
Publisher:
  • Elsevier Sci Ltd, Oxford
Funding / projects:
  • Novel encapsulation and enzyme technologies for designing of new biocatalysts and biologically active compounds targeting enhancement of food quality, safety and competitiveness (RS-46010)
  • Micromechanical criteria of damage and fracture (RS-174004)

DOI: 10.1016/j.matdes.2015.11.088

ISSN: 0261-3069

WoS: 000367235400040

Scopus: 2-s2.0-84953449931
[ Google Scholar ]
43
35
URI
http://TechnoRep.tmf.bg.ac.rs/handle/123456789/3446
Collections
  • Radovi istraživača / Researchers’ publications (TMF)
Institution/Community
Tehnološko-metalurški fakultet
TY  - JOUR
AU  - Dimić, Ivana
AU  - Cvijović-Alagić, Ivana
AU  - Voelker, Bernhard
AU  - Hohenwarter, Anton
AU  - Pippan, Reinhard
AU  - Veljović, Đorđe
AU  - Rakin, Marko
AU  - Bugarski, Branko
PY  - 2016
UR  - http://TechnoRep.tmf.bg.ac.rs/handle/123456789/3446
AB  - Significant enhancement of mechanical properties of metallic biomaterials can be achieved by grain refinement obtained by severe plastic deformation. The purpose of this study was to determine metallic ion release from commercially pure titanium (CPTi) and Ti-13Nb-13Zr alloy processed by high pressure torsion (HPT). The materials microstructures, in the initial state and after HPT deformation, were examined by scanning and transmission electron microscopy. The microhardness was determined along the radius of the disc-shaped samples of ultrafine-grained (UFG) CPTi and Ti-13Nb-13Zr alloy in order to evaluate homogeneity of HPT-processed materials. The quantities of released ions were determined using inductively coupled plasma-mass spectrophotometer for samples immersed in artificial saliva at 37 degrees C for 7 days. Also, the effect of artificial saliva pH value on metallic ion release was estimated. Obtained results revealed that the quantities of released ions from UFG CPTi and Ti-13Nb-13Zr alloy obtained by HPT process were higher than the quantities of released ions from CPTi and Ti-13Nb-13Zr alloy produced by traditional casting. This behavior can be explained by the fact that metallic ions are easily released from microstructure with smaller grains achieved by HPT process.
PB  - Elsevier Sci Ltd, Oxford
T2  - Materials & Design
T1  - Microstructure and metallic ion release of pure titanium and Ti-13Nb-13Zr alloy processed by high pressure torsion
EP  - 347
SP  - 340
VL  - 91
DO  - 10.1016/j.matdes.2015.11.088
ER  - 
@article{
author = "Dimić, Ivana and Cvijović-Alagić, Ivana and Voelker, Bernhard and Hohenwarter, Anton and Pippan, Reinhard and Veljović, Đorđe and Rakin, Marko and Bugarski, Branko",
year = "2016",
abstract = "Significant enhancement of mechanical properties of metallic biomaterials can be achieved by grain refinement obtained by severe plastic deformation. The purpose of this study was to determine metallic ion release from commercially pure titanium (CPTi) and Ti-13Nb-13Zr alloy processed by high pressure torsion (HPT). The materials microstructures, in the initial state and after HPT deformation, were examined by scanning and transmission electron microscopy. The microhardness was determined along the radius of the disc-shaped samples of ultrafine-grained (UFG) CPTi and Ti-13Nb-13Zr alloy in order to evaluate homogeneity of HPT-processed materials. The quantities of released ions were determined using inductively coupled plasma-mass spectrophotometer for samples immersed in artificial saliva at 37 degrees C for 7 days. Also, the effect of artificial saliva pH value on metallic ion release was estimated. Obtained results revealed that the quantities of released ions from UFG CPTi and Ti-13Nb-13Zr alloy obtained by HPT process were higher than the quantities of released ions from CPTi and Ti-13Nb-13Zr alloy produced by traditional casting. This behavior can be explained by the fact that metallic ions are easily released from microstructure with smaller grains achieved by HPT process.",
publisher = "Elsevier Sci Ltd, Oxford",
journal = "Materials & Design",
title = "Microstructure and metallic ion release of pure titanium and Ti-13Nb-13Zr alloy processed by high pressure torsion",
pages = "347-340",
volume = "91",
doi = "10.1016/j.matdes.2015.11.088"
}
Dimić, I., Cvijović-Alagić, I., Voelker, B., Hohenwarter, A., Pippan, R., Veljović, Đ., Rakin, M.,& Bugarski, B.. (2016). Microstructure and metallic ion release of pure titanium and Ti-13Nb-13Zr alloy processed by high pressure torsion. in Materials & Design
Elsevier Sci Ltd, Oxford., 91, 340-347.
https://doi.org/10.1016/j.matdes.2015.11.088
Dimić I, Cvijović-Alagić I, Voelker B, Hohenwarter A, Pippan R, Veljović Đ, Rakin M, Bugarski B. Microstructure and metallic ion release of pure titanium and Ti-13Nb-13Zr alloy processed by high pressure torsion. in Materials & Design. 2016;91:340-347.
doi:10.1016/j.matdes.2015.11.088 .
Dimić, Ivana, Cvijović-Alagić, Ivana, Voelker, Bernhard, Hohenwarter, Anton, Pippan, Reinhard, Veljović, Đorđe, Rakin, Marko, Bugarski, Branko, "Microstructure and metallic ion release of pure titanium and Ti-13Nb-13Zr alloy processed by high pressure torsion" in Materials & Design, 91 (2016):340-347,
https://doi.org/10.1016/j.matdes.2015.11.088 . .