Приказ основних података о документу

dc.creatorMijatović, Nevenka
dc.creatorTerzić, Anja
dc.creatorPezo, Lato
dc.creatorMilicić, Ljiljana
dc.creatorŽivojinović, Dragana
dc.date.accessioned2021-03-10T13:56:59Z
dc.date.available2021-03-10T13:56:59Z
dc.date.issued2019
dc.identifier.issn0584-8547
dc.identifier.urihttp://TechnoRep.tmf.bg.ac.rs/handle/123456789/4097
dc.description.abstractAdjustment and subsequent validation of energy-dispersive X-ray fluorescence (ED-XRF) method for cement based binders with addition of mineral raw materials (fly ash, zeolite and bentonite) was conducted. Eighteen chemical elements present in the material composition were analyzed: ten major elements (Si, Al, Fe, Ca, Mg, S, Na, K, Ti, P) and eight trace elements (Cr, Zn, Cu, As, Ni, Pb, Sr, Mn). Thirty-five samples of either certified reference materials or reference materials of cement, fly ash and clay were utilized during adjustment and optimization of the investigated ED-XRF procedure. The method was consecutively validated in terms of selectivity, precision, working range, linearity, accuracy, robustness, limits of detection and quantification. Thirty-two samples in total, i.e. three certified reference materials and twenty-nine reference materials, were simultaneously analyzed by ED-XRF and inductively coupled plasma optical emission spectrometry (ICP-OES). Comparison of the outputs of monitored methods showed infinitesimally small differences, as correlation coefficients were extremely good (approximate to 1), which highlighted ED-XRF as highly satiable alternative for ICP-OES for the chemical analysis of cement binders.en
dc.publisherPergamon-Elsevier Science Ltd, Oxford
dc.relationinfo:eu-repo/grantAgreement/MESTD/Integrated and Interdisciplinary Research (IIR or III)/45008/RS//
dc.relationinfo:eu-repo/grantAgreement/MESTD/Basic Research (BR or ON)/172057/RS//
dc.relationinfo:eu-repo/grantAgreement/MESTD/Integrated and Interdisciplinary Research (IIR or III)/43009/RS//
dc.relation.isversionofhttp://technorep.tmf.bg.ac.rs/handle/123456789/5039
dc.rightsrestrictedAccess
dc.sourceSpectrochimica Acta Part B-Atomic Spectroscopy
dc.subjectQuantitative analysisen
dc.subjectCoupled plasmaen
dc.subjectFly ashen
dc.subjectBuilding materialsen
dc.subjectMultivariate analysisen
dc.titleValidation of energy-dispersive X-ray fluorescence procedure for determination of major and trace elements present in the cement based compositesen
dc.typearticle
dc.rights.licenseARR
dc.citation.rankM21
dc.citation.spage105729
dc.citation.volume162
dc.description.otherPeer reviewed manuscript: [http://technorep.tmf.bg.ac.rs/handle/123456789/5039]
dc.identifier.doi10.1016/j.sab.2019.105729
dc.identifier.scopus2-s2.0-85075026322
dc.identifier.wos000500481600007
dc.type.versionpublishedVersion


Документи

Thumbnail

Овај документ се појављује у следећим колекцијама

Приказ основних података о документу